hola como puedo ajustar o calibrar un indicador metller toledo gracias


1 Solución propuesta

SOLUCIÓN ACEPTADA
Hello! Adjust you the ID5 balance in the you remove the load cell plug. Thereafter, the ID5 PLUG IN shows in the display. now you hold the reset button down until SERVICE appears. switch ID5, pure screw load cell plug, turn on ID5, it reappears SERVICE in the display. Below is a YES (ENTER KEY), NO (TARE key) Game Now press the ENTER key, it appears RETURN while now with no reply until CAL appears. CAL EXT = YES SET PRELOAD = YES SET FULLCAP = YES UNLOAD = YES parameter = YES Save